A.4.3 Contrast Bolus Module
Attribute Name Tag Type Notes
Contrast/Bolus Agent (0018,0010) 2
Sent if contrast exam, as
entered in user interface.
Contrast/Bolus Route (0018,1040) 3
Sent if contrast exam, as
entered in user interface.
A.4.4 CT Image Module
Attribute Name Tag Type Notes
Image Type (0008,0008) 1
Always sent. See section
2.5.1
Samples per Pixel (0028,0002) 1 Always sent with value = 1
Photometric Interpretation (0028,0004) 1 Always sent
Bits Allocated (0028,0100) 1
Always sent with value =
16
Bits Stored (0028,0101) 1
Always sent with value =
16
High Bit (0028,0102) 1
Always sent with value =
15
Rescale Intercept (0028,1052) 1 Always sent
Rescale Slope (0028,1053 1 Always sent
Rescale Type (0028,1054) 1C
Always sent with value =
HU
KV (0018,0060) 2 Value always sent.
Acquisition Number (0020,0012) 2 Value always sent.
Scan Options (0018,0022) 3 Value always sent.
Data Collection Diameter (0018,0090) 3 Value always sent.
Reconstruction Diameter (0018,1100) 3
Value sent for all images
except scouts.
Distance Source to Detector (0018,1110) 3 Value always sent.
Distance Source to Patient (0018,1111) 3 Value always sent.
Gantry / Detector Tilt (0018,1120) 3 Value always sent.
Table Height (0018,1130) 3 Value always sent.
Rotation Direction (0018,1140) 3 Not sent for scout or axial.
Exposure Time (0018,1150) 3 Value always sent.
X-Ray Tube Current (0018,1151) 3 Value always sent.
Exposure (0018,1152) 3
Value always sent. (See
Note below)
Filter Type (0018,1160) 3 Sent.
Defined terms: